Specifications
Selection
| Model | 6630-1 | 6630-3 | 6630-5 | 6630-10 | 6630-20 | 6630-30 | 6630-50 |
| Frequency | 10Hz-
1MHz |
10Hz-
3MHz |
10Hz-
5MHz |
10Hz-
10MHz |
10Hz-
20MHz |
10Hz-
30MHz |
10Hz- 50MHz |
SPEC
| Model | 6630 | |
| Frequency Range | 10Hz-50MHz | |
| Frequency Setting | Continuity | |
| Frequency Resolution | ±0.01% | |
| Frequency Output Accuracy | 100mHz, 6-bit Frequency Input | |
| Basic Accuracy | typical ±0.05% | |
| AC Drive Level | Test Signal Voltage Level | 10mV – 2Vrms(100Ω)
10mV – 1Vrms(25Ω) |
| Voltage Minimum Resolution | 1mV | |
| Accuracy | ALC ON:6% * Voltage ±2mV
ALC OFF:10% * Voltage ±2mV |
|
| Test Signal Current Level | 200µA – 20mArms | |
| Current Minimum Resolution | 10µA | |
| Accuracy | ALC ON:6% * Current ± 20µA
ALC OFF:10% * Current ± 20µA |
|
| DC Drive Level | 1V (fixed) | |
| Output Impedance | 25Ω/100Ω (switchable) | |
| Measurement Mode | Meter mode, list mode | |
| Measurement Parameters | |Z|、|Y|、θ、X、R、G、B、L、D、Q、DCR、C、ε and μr | |
| Test Time (Fastest) | < 3mS | |
Measurement Parameters
| Parameters | Range |
| |Z| | 0.000 mΩ to 9999.99 MΩ |
| R, X | ± 0.000 mΩ to 9999.99 MΩ |
| |Y| | 0.00000 μS to 999.999 kS |
| G, B | ± 0.00000 μS to 999.999 kS |
| θRAD | ± 0.00000 to 3.14159 |
| θDEG | ± 0.000° to 180.000° |
| Cs, Cp | ± 0.00000 pF to 9999.99 F |
| Ls, Lp | ± 0.00 nH to 9999.99 kH |
| D | 0.00000 to 9999.99 |
| Q | 0.00 to 9999.99 |
| Δ | ± 0.00% to 9999.99% |
| Rdc | 0.00 mΩ to 99.9999 MΩ |
| εr’ εr” | 0 to 100000 |
| μr’ μr” | 0 to 100000 |
System
| Measurement Circuit | Series/Parallel |
| Correction | Open Circuit/Short Circuit/Load correction |
| Cable Compensation | 0 / 0.5 / 1 /2 (m) |
| BIN | ABS、ΔABS、Δ%、OFF |
| Comparator | 9 |
| List Mode | 50 groups of Multi-steps setting (Each group contains up to 15 steps) |
| Bulit-in Storge | 100 sets LCR setting documents, 50 groups of list mode setting |
| USB Host Storge | LCR setting documents, list mode setting document, BMP graphics, Sweep screen and test result data |
| Trigger Test | Auto, manual, RS-232, GPIB, Handler |
| Power Supply | Voltage:90-264Vac |
| Frequency:47-63Hz | |
| Power consumption | 30VA |
| Environment | Temperature:10℃-40℃ |
| Humidity:20-90%RH | |
| Dimension | 336x147x340 mm (W*H*D) |
| Weight | 3 Kg |
| Interface | RS-232、Handler、LAN、USB Host、USB Device |
| Display | Color Screen, 7″ TFT (800*480) |
Application
Automatic Production Line
Connect 6630 RS-232, LAN, GPIB or Handler to automatic machine (measuring speed<3mS),automated production is the most efficiency.
The precision impedance analyzer 6632 has seven frequency options(10Hz to 1/3/5/10/20/30/50MHz).
List Mode
50 groups of list mode setting (Each group contains up to 15 steps).
Four Parameters Displayed Simultaneously
7.0”TFT, 800×480 color screen, and four parameters can displayed simultaneously.
Measure Inductor (DCR)
Inductor coil turns and winding materials will affect DCR value. The key parameters for inductor are Ls / Lp / Q / SRF / DCR.

Magnetic Material Test Fixture (FX-0000C8)
Using the magnetic material test fixture for measuring of permeability of various toroidal cores or ferrite cores and electromagnetic shielding coating materials, 6630 built-in formula to directly calculate the permeability coefficient value μr
- Dielectric Material Test Fixture (FX-0000C7)
- Liquid Dielectric Material Test Fixture (FX-000C20)
Using the dielectric material test fixture for measuring PCB board or ceramic board.
6630 built-in formula to directly calculate the material dielectric value εr ‘/εr ”
Using the liquid dielectric material test fixture for measuring the characteristics of electrochemical materials.